The Intermediate Check: What to Do Between Two Calibrations

The Intermediate Check: What to Do Between Two Calibrations

Good quality IEC 60065 Standard for sales
Good quality IEC 60065 Standard for sales
Our goods has safely arrived at our residence. Thank you for your excellent services. I hope we can cooperate again in the future.

—— Wira

I'm Online Chat Now
The Intermediate Check: What to Do Between Two Calibrations

An annual calibration certificate is a snapshot. Between two snapshots an instrument can drift, be dropped, be repaired, or be used outside its range, and nothing in the file records it until the next certificate contradicts the last one.

The intermediate check closes that gap. It is not a calibration: it does not need a full set of points, and it does not need an accredited provider. It needs a stable reference and a fixed procedure.

Choosing what to check

InstrumentReference usedTypical intervalAcceptance
Hipot testerResistive load or reference dividerMonthlyWithin 2 % of set voltage
Insulation resistance testerReference resistors, 1 MΩ to 100 MΩMonthlyWithin 5 % at each decade
Force gaugeReference load cell or dead weightsMonthlyWithin 1 % of reading
Torque wrenchTorque transducerQuarterly or 5,000 cyclesWithin 4 %
Temperature loggerIce point and a reference thermometerQuarterlyWithin 0.5 K
Glow-wire testerReference thermocouple and 960 °C checkQuarterlyWithin 10 K

The reference does not have to be better than the instrument in every respect; it has to be stable. A 1 MΩ resistor with a 0.1 % drift per year is a better reference than a calibrated instrument that nobody can find.

What makes a check defensible

  • A written procedure with the pass limit on it. A check without a limit is an observation.
  • One point at the working value. Checking a 5 kV tester at 500 V proves nothing about the range in use.
  • A record of the result, not of the activity. "Checked" is not a result; the number is.
  • An action rule. What happens when the check fails: quarantine, re-calibrate, review results since the last good check.

The traceability question

An intermediate check does not have to be traceable to a national standard, but the reference artefact does, at least to the level the instrument needs. In practice a reference resistor with a manufacturer certificate is enough for a 5 % insulation tester and not enough for a 0.5 % instrument.

Our IEC test equipment range includes reference artefacts for the common checks, and the technical notes give the intervals we see in practice across the 15 stations we maintain.

Standards referenced: ISO/IEC 17025 clause 6.4.13 (intermediate checks), ILAC-G24 for calibration intervals, ISO 10012 for measurement management.

Two artefacts that cover most of the workshop

A 1 MΩ and a 100 MΩ resistor, both with 0.1 percent certificates, cover the insulation testers; a 20 N and a 50 N dead weight cover the probe force rigs. Four artefacts, about 200 dollars, and 80 percent of the monthly checks are covered.

Store them in a drawer that only the quality engineer opens, and record the check result next to the instrument, not in a spreadsheet nobody reads.