Glow-Wire Loop Geometry: Why the Tip Temperature Is Not the Coil Temperature

The glow-wire test specifies a tip temperature and a contact time. Both are easy to state and surprisingly hard to hold, because the element that delivers them is a loop of resistance wire that deforms, oxidises and changes resistance with use.

The geometry that matters

The loop has a specified diameter and a specified length of straight section at the tip. A loop that has been bent out of round concentrates heat differently, and a loop that has lost its straight section contacts the sample over a different area.

FeatureSpecifiedEffect of drift
Loop diameter4 mmContact area and heat transfer change
Tip straight lengthPer standardPressure per unit area changes
Tip temperature550 to 960 °CDirectly changes the test severity
Contact force0.8 to 1.2 NPenetration into soft material
Contact time30 s typicalTotal energy delivered

Why the thermocouple reads low

The thermocouple is welded to the tip and measures the tip. The coil behind it is hotter, because heat is lost at the tip to the air and to the sample. A controller that regulates on the coil temperature therefore holds the tip several tens of degrees below the setpoint.

The standard approach is to calibrate with a thermocouple at the tip and to trim the controller so that the tip reaches the setpoint, accepting that the coil runs hotter.

When to replace the loop

  • Diameter out of round by more than 0.1 mm.
  • Straight section shortened by erosion.
  • Visible necking, which raises local temperature.
  • Tip temperature that cannot be held within ±10 K at the setpoint.

Log the loop change date. A lab that replaces loops on a schedule has fewer arguments about why two runs of the same material gave different results.

Standards referenced: IEC 60695-2-10 for the glow-wire apparatus and test procedure, IEC 60695-2-11 for flammability testing with the glow wire.

The contact force nobody measures

The glow-wire tip is pressed against the specimen with a defined force, usually about 1 N. That force is set by a weight or a spring in the apparatus, and it changes when the arm pivots wear or the weight is replaced with a different one.

A force that is 20 % low reduces penetration into a softening material, which changes both the contact area and the amount of material heated. The tip temperature is still 960 °C; the test is still gentler.

Verifying force quickly

With the wire cold, press the tip onto a balance and read the force at the moment of contact. This takes a minute and catches the common problem of a missing or substituted weight.

Check it monthly and after any maintenance on the carriage.

What the specimen sees

The specimen sees a hot spot of a few square millimetres at 960 °C for 30 s, plus whatever heat conducts away through the material. A material with high thermal conductivity spreads that heat and may not ignite; the same material in a thin section may ignite in 5 s.

This is why the test is run on the actual part, at the actual thickness, not on a standard plaque.

The 30-second test is not a 30-second event

The tip is applied for 30 s, but the specimen continues to char and smoke after withdrawal. The verdict is taken at the end of the application plus the afterflame time, which can add another 30 s or more.

Record both times: application and afterflame. A material that ignites at 25 s and self-extinguishes in 5 s is a different result from one that ignites at 5 s and burns for 60 s.

Ambient draught matters more than most labs admit. A 0.3 m/s air movement across the tip lowers its temperature by 20 to 40 K at the 960 °C setting, which is why the apparatus sits in a enclosure with the front closed during the test.