Force Measurement in 20 N and 50 N Push Tests: Where the Uncertainty Actually Comes From

Force Measurement in 20 N and 50 N Push Tests: Where the Uncertainty Actually Comes From

Good quality IEC 60065 Standard for sales
Good quality IEC 60065 Standard for sales
Our goods has safely arrived at our residence. Thank you for your excellent services. I hope we can cooperate again in the future.

—— Wira

I'm Online Chat Now
Force Measurement in 20 N and 50 N Push Tests: Where the Uncertainty Actually Comes From

Test probes are pushed with 10 N, 20 N or 50 N depending on the clause, and the force has to be right because it decides whether a hinged part opens. The load cell is usually the most accurate part of the setup, which is why an 8 % error is possible on a 0.5 % cell.

The rest of the error comes from four places that no certificate covers.

A typical budget

ComponentContributionNote
Load cell calibration0.5 %The certificate value
Off-axis loading2 % to 5 %Hand-held probes are rarely perpendicular
Rate of application1 % to 2 %Fast pushes read high on a dynamic cell
Contact point1 % to 3 %Pushing 5 mm off the axis adds a moment
Temperature0.2 %Small for a compensated cell
Readout rounding0.5 %A display with 0.1 N steps at 20 N
Combined3 % to 6 %Root sum square, dominated by alignment

The alignment term dominates. A push applied 5 mm off the tip axis on a 50 mm lever produces a bending moment that the cell partially senses, and the reading drifts as the part deflects.

Three practical improvements

  • Guide the probe. A linear bearing and a fixed stop remove most of the alignment error.
  • Apply the force slowly and hold. A steady reading at 3 s is more comparable than a peak at 0.2 s.
  • Calibrate in the direction of use. A cell calibrated in compression reads differently in tension, and a cell calibrated on axis reads differently through a lever.

What to write in the report

The report needs the force, the instrument, the calibration due date and the method: hand-held or guided, peak or steady. Without the method, two labs can disagree by 15 % and both be honest.

Our probe force rigs include the guided carriage and the steady-reading fixture, and the technical notes give the budget for a 20 N hand-held test as measured on our own bench.

Standards referenced: IEC 61032 (probes and forces), ISO/IEC 17025 clause 7.6 (uncertainty), ISO 376 for force instrument calibration.

Getting from 6 percent to 2 percent

The guided carriage is the single biggest improvement: it removes the 2 to 5 percent alignment term. The second is a steady 3 s reading instead of a peak, which removes another 1 to 2 percent. Together they take a 6 percent budget to about 2 percent, without buying a better load cell.

  • Fix the probe axis to the rig, not to the operator hand.
  • Use a stop, so every push ends at the same point.
  • Record steady value and peak, and report the steady one.