Creepage and clearance are both distances between conductive parts. They are measured along different paths, and mixing the two produces a number that looks precise and is wrong.
Clearance is the shortest distance through air between two conductive parts.
Creepage is the shortest distance along the surface of an insulating material between them.
On a flat surface, creepage equals the straight line and the two are the same. On any shaped surface — a groove, a rib, a step — they diverge, and the difference is the whole subject.
| Feature | Clearance | Creepage |
|---|---|---|
| flat surface | straight line | same as straight line |
| groove | straight across the top | down and along, if the groove is wide enough |
| rib | straight across | over the rib, following its profile |
A groove narrower than a defined width (commonly 1 mm, or X mm per the standard in use) is treated as if it were not there for creepage purposes: the creepage path bridges the groove rather than following it.
A groove wider than that width is followed, and the creepage becomes longer than the straight line.
This rule exists because a narrow groove fills with contamination and becomes conductive, so following it would overstate the insulation distance.
A rib is followed for creepage only if its height exceeds a defined value. A low rib is treated as flat, and the creepage is the straight line over it.
The reasoning is the same: a shallow rib does not present a real barrier to a surface film.
Measuring creepage along the straight line. The most common error. On a shaped part the true path is longer, so the straight line under-reports — a conservative error, but it rejects good designs.
Measuring clearance along a surface path. Over-reports the clearance, because the surface path is longer than the air gap. This is the dangerous direction.
Ignoring the groove rule. A narrow groove measured as followed gives an inflated creepage and passes a design that should fail.
Measuring to the wrong point. Creepage is measured between the nearest conductive parts, which is not always the terminals — a screw head or a solder bump may be closer.
Use a flexible scale or a piece of thin tape laid along the surface, then measure the tape. This is more accurate than trying to add up segments by eye.
For clearance, a pin gauge set or a pair of dividers set to the required distance is faster than measuring: if the gauge passes between the parts, the clearance is at least that value.
Photograph the path with a scale in frame. A number without a marked path cannot be checked by anyone else.
Both values, the path followed (marked on a photo or sketch), the groove and rib treatment applied, and the standard clause that defines the limits. Without the path, the number is unverifiable.
IEC60061-3 E27 Lamp cap go gauge and no go gauges
IEC60061 B22 Go No Go Gauge of 7006-11-8 & 7006-10-8
Ipx3 Ipx4 Hand-Held Spray Nozzle
IEC 60529 Ingress Protection Test Probe Kit of Access Probes
Manufacture High Precision IEC60950 Telecom Test Probe